The Drying Time Recorder measures and records drying times of all types of coatings, adhesives, etc. It will record drying times for 1, 6, 12 and 24 hour tests.
Features
More Information
On/Off Button
Press to turn the Quadracycle On or Off. When the On/Off button is pressed and the Green LED is energized the instrument is ready for testing. If the On/Off button is held down for two seconds while the instrument is testing the current test will be terminated and the Red LED will energize.
1 Hour, 6 Hour, 12 Hour and 24 Hour Buttons
When the Green LED is energized the Quadracycle is ready for testing. Select the desired test time by depressing one of the four time cycles. Testing will start immediately after depressing the selected cycle. The time selected and "Testing" will be displayed on the LCD. The Quadracycle will stop upon completion of the selected time cycle and display "Test Complete" on the LCD. If there is a power interruption the Red LED will energize and the LCD will display "Test Interrupted". To reset depress the On/Off Button.
Standards
Meets ASTM D5895, ASTM D1640 and ISO 9117-4
Procedure
Attach the arm and ball assembly to the motor shaft by way of the hex screw. Once tightened to the desired position do not forcefully move the arm and ball position without loosening as this can cause damage to the gearbox assembly.
Set the arm to equilibrium by adjusting the rear counterweight before placing the 12 gram weight on top of the ball shaft.
Apply the coating to your test panel using a Gardco Microm Film Applicator or equivalent wet film applicator.
Immediately place the Drying Time Recorder on your test panel and lower the clean 3/8 in. weighted Teflon stylus gently into position.
Press On/Off Button to turn the Quadracycle On or Off. When the On/Off button is pressed and the Green LED is energized the instrument is ready for testing. (If the On/Off button is held down for two seconds while the instrument is testing the current test will be terminated and the Red LED will energize.)
When the Green LED is energized the Quadracycle is ready for testing. Select the desired test time by depressing one of the four time cycles.
Testing will start immediately after depressing the selected cycle and the stylus will then begin to be drawn along the test panel. The time selected and "Testing" will be displayed on the LCD.
The Quadracycle will stop upon completion of the selected time cycle and display "Test Complete" on the LCD. If there is a power interruption the Red LED will energize and the LCD will display "Test Interrupted". To reset depress the On/Off Button
Evaluate the test panel using the supplied Time Template to determine the stages of drying time.
Place the clear Time Template over the circular area scribed by stylus and note appropriate scale number of minutes or hours required for the film to cure, the character of the track by the 3/8 inch Teflon ball stylus relate to the drying of the film (Reference ASTM D5895).
Record the results and if desired, record the temperature, humidity and the wet film thickness as these factors are important in considering the drying time of films.